System Performance:
- Airflow: High capacity 9 l/s (18 scfm) continuous airflow optimizes temperature
transition rate and throughput.
- Typical Temperature transition rate:
-55° to +125°C: approximately 7 seconds
+125° to -55°C: approximately 20 seconds
- Temperature Control: Air or Patented DUT Dual Loop directly at the DUT case to ±0.1°C
- DUT Sensor Ports: Four: Internal Diode, Type T or Type K thermocouple, and 100 ohm RTD
- Temperature set, display and resolution: ±0.1°C
- Temperature Accuracy: 1.0°C ( calibrated to NIST standard)
- Remote interface ports: Four: IEEE-488, RS232C, SOT/EOT/SFF, Ethernet.
*Note 1: Due to the use of HCFC-free refrigerants with 50 Hz systems,
the ultimate low temperature capacity of 50 Hz systems will be up to 5&C warmer than
60 Hz systems.
Operating Features:
- User Control at Color Touch-screen at front panel or via remote interface
- Two modes of operation:
- Operator Mode: Test and cycle at Hot/Ambient/Cold temperatures
- Cycling Mode: Utilize up to 18 thermal cycling sequences, graphing, datalogging, and
other advanced features.
- User-defined "At Temperature" Windows, airflow rate, Air or DUT Temperature
Control
- Thermal head raising and lowering: Pneumatic control; also manually adjustable
- Arm/manipulator movement: manual locking, Rotates 330 degrees around base for
positioning; extends to 139cm (55 in.)
- Thermal Cycling Sequences: Set up to 18 cycling routines in sequence on one table.
- Hard drive or diskette storage of test set-up files for quick recall and test
repeatability.
- Documentation: Color graphing, Automated datalogging to hard disk and/or diskettes
- Peripheral ports for user-supplied printer, monitor, keyboard, and mouse
- CFC-free for all systems; HCFC-free for all 50 Hz Systems
- CE Approved and SEMI S2-93A compliant

Applications:
- Test and cycle a DUT under a variety of conditions, creating test sets of up to 18
cycling sequences. Graphing, datalogging and printer port features assist in
documentation.
- For 24 hr/ 7 day test environments, the TP04300A's high capacity airflow and DUT thermal
precision ensure high throughput with consistent repeatability. Operator's Mode ensures
quick set-up for testing at up to three temperatures.
- Thermal source for applications requiring precisely controlled thermal air, including
the ThermoFixture® enclosure/ATE interface for thermally testing PCB, modules,
assemblies, digital, high frequency (RF, microwave), and high power devices.

Options & Accessories
Custom Thermal Test Enclosure (CTTE)
Custom Thermal Test Enclosures provide a frost-free environment for testing components,
modules, small subassemblies, printed circuit boards and MCMs over the full -80° to
+225°C temperature range. Patented probe-through windows permit the probing of the PCB or
subassembly while it is held at precise temperature in the CTTE thermal environment, from
-55°C to +125°C.
Click here for Data Sheets / Application
Notes on CTTE.
Flexible Extender
extends the reach of the ThermoStream Thermal head. Available in standard and
"ThermoFixture Quick-Disconnect" configuration.
ThermoWand Assembly
Handheld thermal wand connects to ThermoStream head for testing individual devices on a
PCB or at the tester site. Thermal shrouds of various sizes adapt the ThermoWand for the
DUT.
For other options and accessories (i.e. thermal caps, shrouds and insulation kits), please
contact Temptronic directly.
See TP04300 Brochure and Datasheet
|
 |